Milltronics CVCC Manual de instrucciones Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 112
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
SYNCHRONIZATION
Where two to a maximum of twelve transducers will be sharing a common conduit the
OCM-3s should be synchronized. In order to synchronize OCM-3s:
» remove jumper J1 on board A on all but one OCM-3
» interconnect the SYNC terminal (TB1-20) of all OCM-3s. Insure that
all OCM-3s share a common ground (TB1-34).
18 OCM III 7ML19985AB01